Calibrate SGM
This option lets you calibrate the spherical grating
monochromator. This section describes the specific procedure of
how to do it from SM_GUI. Refer to the beamline
manual5.3 for
background information and exact tabulated absorption peak
positions.
Note that you need access to the current SGM calibration file on
to local file system (mounted from x1a.nsls.bnl.gov,
which is the case when you do the calibration from the microscope
control client computers)5.4. The SGM is
driven by a stepping motor, and the x-ray wavelength (not energy!)
changes linearly with motor position (to a very good
approximation). Therefore, the calibration requires two parameters
(see Eq. 10.2). In daily operation, however, only the
zero_order_steps might change
(steps_per_angstrom should stay fine), and you can
calibrate only this one parameter in SM_GUI. A full
calibration has to be done outside the GUI.
The regular procedure is as follows. Ask the beamline staff for
help in special cases like iron edge etc.
- Record a point spectrum over the appropriate energy
range (280 to 310eV on outboard, 520 to 540eV on inboard,
in steps of 0.1eV each). On inboard the chamber must be
evacuated and backfilled with Helium. Set this spectrum to
by clicking the button in the GUI window. You might want
to save the scan and write down the file name.
- Add the calibration gas by opening the CO
flow on
outboard, or letting some air (Oxygen) leak into the chamber
on inboard (use the valve on the outboard-downstream side of
the chamber). Record another point spectrum over the same
energy range. Again, save the file and remember the scan
number for future reference. Check the "Show
" box.
Identify the calibration peak (e.g., the 292.74eV CO
absorption peak, see the beamline manual for a table of
calibration peaks).
- Fit the peak by dragging the left mouse button over it
(see Fig. 5.19). Don't include any part of
a neighboring peak! A red line will be displayed at the center
of the fitted Gaussian, and the peak position will be shown
below the plot as well as in the display information area. (You
can also just click on the peak rather than fitting).
- If the calibration is off (the fitted peak position
deviates from the tabulated value), click the "Calibrate SGM"
button in the menu. A window will pop up (see
Fig. 5.19). The fitted "wrong" value will
already be entered in the upper field. Into the lower field,
enter the correct, tabulated value. If you click "Accept", you'll be
asked for a confirmation to overwrite the SGM calibration
file. A backup copy of the file will be made automatically.
Note: If you do the calibration from a computer other than the
microscope control client computer, you'll have to specify the
SGM calibration file. Better know what you are doing!
Figure 5.19:
Monochromator calibration
|
![\includegraphics[width=4.in]{images/sm_gui_sgm_cal}](img52.png) |
Holger Fleckenstein
2008-07-08