The Alignment Menu
This menu is inactive in display_only mode. It has the
following entries:
- Read Alignment File
- will make the microscope control
program read the current alignment file.
- Edit Alignment
- will pop up a window (see Fig. 5.12)
which allows you to insert stages and detectors, record certain devices aligned
etc. Chap. 2 will tell you which one to use in which situation.
- Insert/Remove
- Detector: Insert VLM: Moves the Visible Light Microscope into its inserted
position.
- Detector: Insert COUNTER: Moves the counter 2mm downstream of its inserted
position.
- Detector: Insert SIDET: Moves the segmented silicon detector 2mm downstream
of its inserted position.
- Detector: Retract Detector: Moves the detector stage all the way
downstream.
- Detector: Remove Detector: Moves the detector stage to its removed
position (usually downstream/outboard).
- Insert Laser: Moves the laser to its inserted position.
- Insert Photodiode: Moves the photodiode to its inserted position.
- Insert Specimen: Moves the specimen ZSTG into focus position at
the current energy. Also moves ZOSA and ZDET (if detector is inserted)
to their ideal positions.
- Retract Specimen: Moves the specimen to its retracted position (usually 2mm
downstream of the focus position). Also retracts the detector stage. This is for safely
exchanging specimens.
- Insert Specimen Stage: Moves the specimen stage to its retracted position
(2mm downstream of the focus position). This is typically used, if the specimen stage had
been removed first, or if motors have been homed.
- Remove Specimen Stage: Moves the specimen stage to its removed position
(usually downstream/outboard). Since there is not enough space for the specimen stage to
pass by the detector stage in its downstream position, the VLM is moved in first, and the
specimen and detector stage are moved step by step, until removed. This is typically used,
to insert or remove optics (ZP or OSA).
- Record Specimen
- Record Specimen in VLM Focus: Hit this button, when the newly inserted
sample is in VLM focus. This will allow you to move the sample within about 100
m of
its focus position, when inserting it.
- Record Specimen in X-ray Focus: Hit this button, when the specimen is in
X-ray focus. Based on this position, the new focus positions at different energies are
calculated.
- Record Advanced Only use this menu, if you really know, what you are doing!
- Record ZP in VLM focus: After aligning the ZP click this button, when the
VLM is focused onto the ZP.
- Record OSA in VLM focus: After aligning the OSA click this button, when
the VLM is focus onto the OSA.
- Record Detector Aligned: COUNTER: Click this button to use the current
detector positions as the aligned positions for the counter.
- Record Detector Aligned: SIDET: Click this button to use the current
detector positions as the aligned positions for the segmented silicon detector.
- Record Detector Aligned: SIDET: Click this button to use the current
detector positions as the aligned positions for the segmented silicon detector.
- Drift: Set Current as Image1: Sets the currently open image
as image #1 for calculating the drift parameters (see
12.3).
- Drift: Set Current as Image2: Sets the currently open image
as image #2 for calculating the drift parameters (see
12.3).
- Drift: Calculate Parameters: Calculates the new drift
parameters from the two previously set images (see
12.3).
- Drift: Record Parameters: Click this button, to use the entered or calculated
parameters (in the fields below) as the new drift parameters. This is to activate software
correction for misalignment of the microscope Z-Axis with respect to the X-ray beam axis
(see Sec 12.3). Recording 0.0 for both parameters
will disable drift correction.
- You can enter the drift parameters (dx_over_dz dy_over_dz) by hand, if you want.
- By checking one of the boxes Sobel or Roberts you can use one
of the algorithms for edge enhancement, before calculating the drift in the two images.
This may produce slightly different results. Experiment will show, which of them is the
best one to use.
- Sidet Z-Matching Parameter: This
parameter describes the Z alignment of the silicon
detector. It is defined as the ratio of the beam cone on
the detector chip to the diameter of the chip's quadrant
segments. A value of 1.0 therefore means that the beam
cone exactly fills the quadrant segments, this is
"standard". The value has to be entered by hand when the
detector is aligned. The value is transferred to the
server (to be written to the data files) as soon as you
enter it into the field and press Enter or move the
keyboard focus out of the field. The value is important
for quantitative reconstruction of phase.
- Align and Focus Sample with VLM
- This button will pop
up a window to move the three motors which are being used to
pre-align and focus the sample with the VLM: The X and Y
sample stepper stage and the Z detector stage (see
Fig. 5.13). You could also do that with
the stage and detector motor-move windows described in
Sec. 5.13, but here the required motors
are conveniently combined in one window, without the danger of
accidentally moving the VLM for X and Y alignment or the
sample for focusing.
Figure 5.13:
The popup window to align and focus the sample
with the VLM
|
![\includegraphics[width=2.in]{images/sm_gui_vlm_align}](img44.png) |
- Write Alignment File
- will save all current alignment
parameters to the alignment file.
Holger Fleckenstein
2008-07-08