Venting from helium to atmosphere
- Open the helium gas bottle.
- Close V9i for safety.
- Loosen the locking nut on one viewport to allow it to pop open when at atmosphere.
- Press the Vent Over button and hold it until the
chamber reached atmosphere. You hear when the viewport pops
open. Release the button to stop the flow of helium.
- Remember to close the helium gas bottle at the end of the day.