The silicon detector calibration file

The configuration files for the segmented silicon detector sidet_calibration_stxm[i,o].dat are used in converting the raw data (volts) for each detector segment into calibrated data (kHz). Basically, all the values will be copied into each recorded data file.

Some of the values have to be determined only once for each new detector installed, by following the instructions about detector calibration in the silicon detector manual13.1. The IDL program detcal_gui can write them directly into a calibration file (which has to be copied manually to the appropriate location), or they can be entered into the file by hand. Others have to be updated frequently by re-calibrating the detector with the appropriate button in the microscope control program (because the leakage current in the silicon chip can change due to ambient temperature variations). After you have changed the former ones, you will have to restart sm_server, to make them become activated.

For more information about the silicon detector calibration, see Sec. 10.7 or the silicon detector manual at http://xray1.physics.sunysb.edu/user/manuals.php.

The values are (parameter example):

Holger Fleckenstein 2008-07-08