The easier way is to let software correct for the drift. In this option at each energy change the
piezo stage moves the sample in
and/or
a little (according to the entered drift
parameters) to keep the same spot of the sample in the focus position.
For this just enter the previously calculated drift parameters in Alignment
Edit Alignment
Record Advanced and click Drift: Record
Parameters. You may want to Write the Alignment File to prevent the parameters form
getting lost at a computer crash.
Disadvantages of this method:
Holger Fleckenstein 2008-07-08