The next step is to get beam into the microscope chamber. Go to some relatively large beamline
slit settings like 200/200/0 slits, for ENS_I=200
m, EXS_I=200
m,
and EXSY_I wide open. Look in the upstream inboard viewport; you should see the snout
as a long metal tube with a glass section in its middle (see Fig. 12.2). Inside
this glass section is a mirror that lets you look down the direction of the beam path towards the
exit window between the beamline vacuum and the inside of the STXM chamber. You should be able to
see beam on the phosphor around the exit window, and adjust the chamber accordingly. Of course you
have to have all valves and shutters in the beamline open, for seeing beam on the phosphor.
|
If the microscope is close to being aligned, you can simply monitor the flux on a detector to
evaluate the alignment of the microscope chamber to the beam. If you are really starting from the
beginning, you will need to remove the zone plate and OSA (see Sec.
12.2.4), and scan a 5-10
m diameter pinhole as the sample to evaluate
the illumination of the exit window. If the exit window is not illuminated homogenously, you will
have to move the the whole STXM chamber by turning the manipulators which control the tripod it is
sitting on. Do iterative steps of moving the chamber and scanning the pinhole over the exit
window, until you are satisfied with its illumination. At the end of this iterative
procedure, you should have reasonably even and bright illumination of the exit window.
In between scans you always need to make sure, you still get signal into the detector. You can
even scan the detector in (XDET vs. YDET) and center it according to that scan. (You can make sure
certain detectors - not the proportional counter though - are working by seeing if they respond to
visible light leaking into the chamber). Whenever you change the alignment of the detector don't
forget to record the new alignment.
Holger Fleckenstein 2008-07-08