Dark signal subtraction and conversion to photon count rate

The x-ray flux (photons per second) into one detector segment is calculated as follows:

\begin{displaymath}
\Phi = \frac{F}{C} \frac{U_{\mbox{out}} - U_0 - C \times
I...
... \times t_{\mbox{dwell}}}{t_{\mbox{dwell}} -
t_{\mbox{dead}}}
\end{displaymath} (10.4)

[1, Eq. 3.9], where

The calibrated signal (count rate) is calculated from the raw data automatically by SM_GUI or read_stxm5.pro. Make sure you look at the correct signal (see Chap. 9)! If you want to do it yourself, apply the two formulas listed above.