Next: Power filter box
Up: Detector Installation
Previous: Components
Contents
Fig. 4 shows the front face of the main power
supply (Kenwood). The tracking is turned on (positive and negative
supplies have the same settings). The supply voltage is +/- 12V
and the current is limited to 0.15 A. The output protect should be
turned on to limit the current if something goes wrong. The output
has to be turned on with the Output button on the upper
right. The red light over the button will turn on. The power
supply has a bridge that connects the common and
ground terminals on the output, which has been disconnected
lately.
Figure 4:
Kenwood power supply
![\includegraphics[width=0.5\columnwidth]{images/kenwood}](img6.png) |
Microscope User