Next: Interfacing Of The Detector
Up: Detector Alignment
Previous: X and Y Alignment
Contents
To align the detector in Z, you have two options. Note that you
probably won't find the "perfect" position, but that's ok.
- Scan XDET or YDET vs.
ZDET. Since the big X detector stage doesn't move
very smoothly, you might be better off with YDET.
Again, there should be no sample inserted, or you should be
inside a pinhole. Scan, say,
pixels with a
pixel size of 3 microns in X and 5 microns in Z. Display the
signal from segment 8 and use the "Stern Special" colortable
again (see Fig. 21). The ideal Z position
is around the tip of the black inverted triangle. Also look at
the horizontal profile through the image - at the correct Z
position, the profile should just lift off from zero at the
center.
Figure 21:
YDET vs. ZDET scan for detector
alignment. Only the signal from segment 8 (the outer ring) is
displayed, and the Stern Special colortable is used. Move the
detector to the tip of the black triangle, where the
horizontal profile just touches zero.
|
|
The file x1ais_20051028_0039.sm is an example for
such a scan.
- Insert a pinhole as sample and focus. Go to the center
of the pinhole and then scan XPZT or
YPZT vs. ZDET. Look at the signal from
segment 8 again (see Fig. 22). The two
vertical lines are more or less constant in intensity, that's
scattered light from the pinhole edges. Outside the pinhole
should be constantly dark. Inside the pinhole, the intensity
should go from high (detector too far) to low (detector too
close). The ideal Z position is where the intensity inside the
pinhole (look at the Y profile) just starts to lift up from
the constant base.
Figure 22:
XPZT vs. YDET scan with a 10 micron
pinhole installed. Only the signal from segment 8 (the outer
ring) is displayed. The correct Z position for the detector is
just where you start getting some signal inside the pinhole.
|
|
The file x1ais_20051028_0040.sm is an example for
such a scan.
Note that the Z alignment is really not that critical. For an
80micron diameter zone plate with a focal length of 1mm, the
ideal detector position is
from the focus. If you are off by 0.5mm in Z
(which is a lot), the diameter of the beam cone changes by
in diameter or
20microns in radius. That's 20 percent of the width of the
quadrant segments. And with the methods described above, you
should find the best Z position better than 0.5mm.
Next: Interfacing Of The Detector
Up: Detector Alignment
Previous: X and Y Alignment
Contents
Microscope User
2007-02-02