Next: Correction For Detector Dark
Up: Old Notes about STXM 4
Previous: Old Notes about STXM 4
Contents
For background on the crosstalk, see Michael Feser's thesis. The
crosstalk is determined by raster scanning the detector through a
collimated beam (10
m diameter pinhole downstream of exit
window). An image of the detector segments is obtained. In the
following, we will denote the detector segments by the number of
the software channel which reads it out (see
Fig. 18). Refer to table
6 to see which software channel
corresponds to which detector segment, as given by the chip
design.
Table 6:
Software channels and corresponding detector segments,
as given by the chip design
| Software Channel |
Detector Segment |
| 0 |
1d |
| 1 |
2d |
| 2 |
3d |
| 3 |
4d |
| 4 |
5d |
| 5 |
6d |
| 6 |
7d |
| 7 |
8d |
|
Figure 18:
Image of the detector segments, obtained by a detector
scan with a pinhole installed. The numbers correspond to the ADC
channels. The image shows the detector chip turned by 90 degrees
because for detector scans, the Y axis is the fast axis (the
chip is actually installed with the three inner segments
horizontally).
|
|
For each segment
, you will have to correct the
crosstalk with all other segments
(except
).
Refer to these indices in the following description.
Start the detector calibration routine with the keyword
crosstalk set:
IDL> det_cal, /crosstalk
- Select the microscope you are using (sxtmi,
stxmo or cryo).
- Select the detector map file that you just recorded in the
popup.
- Another popup window will display the data channel (ADC channel)
. This will be the first segment for which you correct the
crosstalk with the other segments
. It will be clearly visible,
and any other segments
"crosstalking" to it will be faintly
visible.
- By typing m, you can change the display maximum to
any value between 0 and 100%, where lower values will show the
crosstalking channels more clearly.
- For all segments
crosstalking to segment 0,
do the following:
- Type the number of the channel (segment) you want to
correct, let's say
.
- Then give a guess for the crosstalk in per cent. If the
crosstalking segment is brighter than the surrounding, you will
have to give a negative number, and vice versa.
- Repeat the above until the crosstalking segment disappears.
You have to type in the channel number each time.
- By typing m, you can change the display maximum
setting at any time, to make slight brightness differences more
clearly visible.
- Do the correction for all other channels
by just typing
the corresponding channel number and playing around with the
crosstalk percentage.
- Then type x to exit. The program will jump to the
next channel
.
- For all other segments
, do the same.
Next: Correction For Detector Dark
Up: Old Notes about STXM 4
Previous: Old Notes about STXM 4
Contents
Microscope User
2007-02-02