To do the detector calibration properly, you should understand the
background as described in Michael's thesis and the next chapters.
However, if you want to get up to speed quickly without reading
everything right now, use the following guidelines to go through
the calibration:
- Record a detector map file as described in
Sec. 6.3.
- Record four or five dark scans as described in
Sec. 6.3.
- Start the calibration program in IDL and specify the
number of segments (8 is default):
IDL> detcal_gui, sidet_n_segments=8
- Read the detector map file from "File" / "Read Detector
Map File".
- Do the crosstalk correction as described in
Sec. 6.5.
- Click the "Dark Current Correction" button and select
all the dark scans you took above.
- If you want, click the "Dark Current Histogram Analysis"
and "Dark Current Fourier Analysis" buttons. The program will
create a number of files as noted in
Sec. 6.8. Check them if
they make sense, or to identify noise.
- Make sure the correct values for the detector dead time
and eV per Femtocoulomb are entered in the two fields. The
dead time is the delay plus width of the S2 pulse setting on
the Quantum Composers timing module and should be 0.101ms.
eV per Femtocoulomb should be 22594.25. Press Enter in the
field after changing any of them.
- Load an initial setting for the C tune values from a C
tune file or an existing calibration file. In the latter case,
make sure you don't overwrite the other parameters previously
obtained by checking only the
mvolts_per_femtocoulomb checkbox when reading the
calibration file.
- Click the "C Tuning" button and follow the instructions.
For more information, see Sec. 6.9.
- Write out the detector calibration file ("File" / "Write
Calibration File"). Name it appropriately and copy it to the
correct location so that the microscope control program can
find it (e.g.,
/mnt/x1a/sidet_calibration_stxmi.dat).