To bring the beamline back from a shutdown position, these are some
of the things to do:
- Make sure that air pressure, water supply and return pressure
are all non-zero (gauges underneath V8L). Check that flow to LBL mirror
reads over 1 gallon/minute (digital gauge under LBL tank)
- Make sure you have an approved safety form posted. (If you had
it pre-approved, it is in the hands of the safety operator. If not, you are
out of luck).
- Check all the vacuum gauges, to make sure that the vacuum is OK,
especially if some of the valves got closed. (Compare with weekly status
reports). Make sure ion pumps are running. If all OK, open valves.
- Fill out safety check list, and ask the operations coordinator
to open up
the front end. (Blank forms are in a sleeve below the ``Green board,''
and they're also available for printing from a
separate web page.)
- Make sure undulator gap is set right. If not, request
the proper gap from the control room.
- Verify entrance and exit slit sizes, wavelength setting, and
photodiode reading. Open safety shutter.
- If there is signal at PD
, but not PD
, either the shutter is closed, or the
differential pumping section got bumped, or the LWB
mirror is set wrong. Check beam on phosphor at
upstream side of exit window.
- If there is signal at PD
, but no counts in
PC, check the alignment of the snout, PC HV, and gas
flow to counter. Counter may also be misaligned, or
sample may be blocking the beam.
- Check to see that feedback is on. (If not, contact control room).
Holger Fleckenstein
2008-07-08