Energy resolution

The calculated resolving power of the inboard (X1A2) monochromator is shown in Figs. 3.11 and 3.12.

Figure 3.11: Resolving power (top), energy spread (middle), and proper exit slit width for single-mode illumination (bottom) of the inboard, X1A2 branch SGM. This is for the ``standard'' exit slits EXS_I and EXSY_I at a size indicated in the bottom figure, and entrance slit sizes as indicated on the top and middle plots. Calculated using the routine sgm_delta_ev.pro.
\includegraphics[width=5.in]{sgm_inboard}

Figure 3.12: Resolving power (top), energy spread (middle), and proper exit slit width for single-mode illumination (bottom) of the inboard, X1A2 branch SGM. This is for the ``high energy'' exit slits EXS_IH and EXSY_IH at a size indicated in the bottom figure, and entrance slit sizes as indicated on the top and middle plots. Calculated using the routine sgm_delta_ev.pro.
\includegraphics[width=5.in]{sgm_highenergy_inboard}